Troubleshooting
Problems
Compressor overheats.
The compressor does not operate.
Medication does not nebulize properly.
The device is noisy.
Should the device still not operate properly after checking the unit as indicated
above, contact Respironics Customer Service at 1-800-345-6443 or 1-724-387-4000.
Solutions
Turn the unit OFF for a minimum of 40
minutes to allow the compressor to cool.
Check all power sources to the compressor.
Push the ON/OFF button again.
If you checked all the items above and
the unit still does not work, allow the
compressor to cool down for a minimum
of 40 minutes before turning the
compressor on again.
Check for proper connection of the
tubing adapter.
Make sure the nebulizer is not empty.
Clean the nebulizer as described in
the manual.
Replace the nebulizer.
Check the intake filter and replace it if it
is dirty.
Check that the filter cover is correctly
positioned.
Specifications:
Power Supply:
Weight:
Size:
Maximum pressure:
Maximum flow rate:
Mode of operation
Class II Device (Double Safety Insulation).
Type B Device (device with specific protection against electrical hazards).
Certification:
Electric Safety Standards EN60601-1
Respiratory therapy equipment. Nebulizing systems and their components EN13544-1
Electromagnetic Compatibility according to EN60601-1-2
Storage Conditions:
Temperature:
Humidity:
Operating Conditions:
Temperature:
Humidity:
Replacement Parts and Accessories:
Filter
1041433
AC power/battery charger
1041517
DC Car Adapter (12V input/8.5V output)
1041515
Rechargeable Lithium Polymer Battery
RDD499
Pediatric Face Mask
1048154
Adult Face Mask
1048285
Micro Plus Nebulizer
1041431
Symbol Glossary:
Attention: Read accompanying documents
~
Alternating Current
Optional Rechargeable Battery:
Operating time with the battery
Up to 60 minutes
Number of treatments per charge
Up to 4 treatments when nebulizing 3 ml of medication
Normal time to charge
Approximately 3.0 hours
Charge cycle
Approximately 400 cycles
Electromagnetic Information:
Portable and RF communication devices such as cellular phones, pagers, etc. can interrupt operation of electrical
medical equipment. For this reason, your compressor must be placed far enough away from these devices to prevent
interference. This device meets IEC60601-1-2 standard for electromagnetic compatability (EMC).
EMC data sheets (1047862) are available upon request from Respironics, Inc. customer service at
1-800-345-6643 or 1-724-387-4000 or electronically at www.microelite.respironics.com
8.5VDC, 1.4A Max
1.0 lbs. (with battery)
0.66 lbs. (without battery)
3" x 2-5/8" x 1-7/8" (without battery)
3" x 2-5/8" x 2-1/2" (with battery)
17 PSI, 120 KPa
5 LPM
20 minutes ON/40 minutes OFF
MIN -4°F (-20°C) – MAX 140°F (60°C)
MIN 10%RH – MAX 95%RH
MIN 41°F (5°C) – MAX 104°F (40°C) (with tubing)
MIN 41°F (5°C) – MAX 77°F (25°C) (handheld)
MIN 10%RH – MAX 95%RH
ON/OFF (power)