Technical Specifications

Temperature measurement range: 32.0°C – 42.2°C (89.6°F – 108.0°F)
Operating temperature range:
Storage temperature range: -
• Ear Thermometer meets accuracy requirements established in ASTM Standards
E1965-98.
• Batteries: two AAA batteries.
• A clinical summary is available upon request.
• For home use only.
Temperature
ASTM laboratory accuracy requirements in the display range of 36°C to 39°C
(98.8°F to 102.2°F) for infrared thermometers is ±0.2°C (±0.4°F), whereas for
mercury-in-glass and electronic thermometers, the requirement per ASTM
Standards E667-86 and E1112-00 is ±0.1°C (±0.2°F).
The Accu-Light 1-Second Ear Thermometer has an operating range of 5.0- 40C. The
maximum error applies to the operating range 16-40C as specified in thee ASTM
E1965-98.
10
All manuals and user guides at all-guides.com
Maximum Laboratory Error
Celsius Scale:
36˚C to 39°C
Outside this range
Fahrenheit Scale:
96.8˚F to 102.2°F
Outside this range
5.0˚C - 40.0˚C (41.0˚F - 104.0˚F)
-20˚C - 50˚C (4.0˚F - 122.0˚F)
Maximum Error
±0.2°C
±0.3°C
±0.4°F
±0.5°F