VitalStim® Plus Electrotherapy System
Lead wires
Lead wire Clips (attached to lead
wire)
Operator Remote Switch
OPERATOR REMOTE SWITCH
To operate the Patient Remote Switch, plug the remote into the device
on the Jack Panel, as shown below:
After connecting the electrodes and setting up VitalStim or VMS,
complete the following steps to activate Operator Remote Switch
(activated switch indicated by blue icon):
1 .
Connect Remote Control
2 .
Select the stimulation channel and adjust intensity to desired
level
3 .
Press and Release Remote Control button to activate Manual
Mode . Intensity will decrease to 0 mA
4 .
To start stimulation, press and hold the Remote Control button
5 .
To stop stimulation, release the button
6 .
To adjust intensity level, press and hold the remote control
button while increasing or decreasing intensity .
CAUTION
•
Operator Remote Switch to be used under supervision of a
physician or certified VitalStim user only .
THERAPY SYSTEM START-UP
Complete the following steps for initial setup of the VitalStim® Plus
Electrotherapy System:
1 .
Remove battery cover, insert batteries, place back the cover .
2 .
Press the ON/OFF button located on the front of the device:
3 .
Select desired function on the Home Screen (shown below) .
5
GENERAL INFORMATION
SCREEN DESCRIPTION
Channel Area
Located at the bottom of each screen, this screen displays the
following status information about each channel:
n/a: Indicates the channel is not (yet) available to be selected
Available: Indicates the channel is available for use
Running: Indicates a treatment for the channel is currently
running
Paused: Indicates a treatment is currently paused
No contact: indicates open circuit which could be caused
due to poor electrode contact or fault with lead wires being
damaged or not connected properly
The image below shows the Home screen with modality and resource
icons .
Title
Bar
Main
Area
Channels