Transmission Losses; Problems Inside The Usb-Receiver - AREX TL-300 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

1. Transmission losses

Sometimes transmission losses may arise, indicated by missing
temperature data in the sensor's curve display.
Data losses may be caused by:
1.

Problems inside the USB-receiver

2.
Problems in the temperature sensor module
3.
Problems in the signal transfer between temperature sensor
module and USB-receiver
1.1. Problems inside the USB-receiver
The receiver does not registrate a single data signal, even if the sensor
is located at a minimum distance to the receiver.
Potential problems:
-
USB-cable between receiver and computer is missing or defect
-
Improper installation of the USB-module
-
Unknown software problem in the computer system
Suggested solutions:
-
Check the display window in the temperature logger display
for a field in the lower left area. The field is to display the value
'ready' continously. If the display intermittently displays
'RF_USB-Communication failure', the Windows operating
system failed to find the USB-module.
-
Remove the USB-cable, wait about ten seconds and reconnect
the cable.
-
Deinstall the temperature logger application software and
reinstall it again.

E
N
G
L
I
S
H

Publicité

Table des Matières
loading

Produits Connexes pour AREX TL-300

Ce manuel est également adapté pour:

Tl-500

Table des Matières