Key Functions / Description:

1. ON/OFF:
Press to turn scale ON or OFF.
2. TARE/ZERO:
seconds accesses application menu.

3. "F" Key:

Function key for applications (Unit toggling, Counting & Totaling ) and menu scrolling.
4. CAL/PRINT:
Press and "hold" CAL/PRINT key (for approximately 3 seconds) and scale will automatically enter
calibration mode (displaying one of three weight values). Press and "release" CAL/PRINT key to print
displayed information when utilizing RS-232 interface.
5. LIGHT :
Turns on backlight during battery operation (Auto Off: One Minute).

Calibration:

1
Weight Calibration:
"CAL/PRINT" key for more than 3 seconds to enter the calibration mode.
Depending upon your model a default weight will be displayed in kg (note; you have 3 weight values
2
to choose from).
3
In order to select additional cal-weight options, press and release the "F" key and the display will
toggle between selectable values based on model. (Please see Product Specifications)
4
After selection of the desired cal-weight option, press the "CAL/PRINT" key and release to confirm,
and the value will flash to begin calibration.
Place corresponding weight onto center of pan. If weight is accepted, the display will stop flashing and
5
show "selected kg" value in display .