Ensemble de transducteurs ultrasons

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 5
4. Additionally required equipment
1 Function generator, 130 kHz (230 V, 50/60 Hz)
or
1 Function generator, 130 kHz (115 V, 50/60 Hz)
1 Analog oscilloscope, 2x30 MHz
3 Stand bases, 0.5 kg
2 HF cables, BNC/4-mm plugs
2 Adapters, BNC/4-mm safety sockets
5. Operation
5.1 Calibration of resonant frequency
Set up the ultrasonic transmitter and receiver
close together and facing one another.
Connect the transmitter to the 50 Ω/600 Ω output
of the function generator and set the frequency to
40 kHz.
U21010
U21010
Fig. 2 Experiment set-up for the reflection of ultrasonic waves by a projection screen
3B Scientific GmbH • Rudorffweg 8 • 21031 Hamburg • Germany •
U21010-230
U21010-115
U11175
U8611210
U11257
U13807
S
Fig. 1 Experiment set-up for calibrating resonant frequency
S
R
Subject to technical amendments
© Copyright 2010 3B Scientific GmbH
Connect the receiver to an oscilloscope.
Observe the received signal and make fine
adjustments to the frequency to maximise the
signal amplitude.
5.2 Example
Set up the ultrasonic transmitter and receiver
alongside one another in front of the projection
screen.
Connect the transmitter to the 50 Ω/600 Ω output
of the function generator and set the frequency to
the resonant frequency (see 5.1).
Connect the receiver to the 30-W amplifier in the
function generator.
Connect the output of the amplifier to the
oscilloscope.
Move the projection screen and observe the phase
differences between the signals.