PESOLA MS500 (500g · d = 0.1g) Manual
I. General conditions for use
Please carefully read the following instructions before use:
• Handle with care. Do not attempt to pull up the platform. Gently place the load
onto the middle of the platform. Rough handling such as shaking, dropping or
otherwise shocking the scale may permanently damage the sensor. Loss of
warranty will result.
• Avoid any exposure to extreme heat or cold. Your scale works better in normal
room temperature. After great temperature fluctuation allow the scale to acclimate
for one hour – particularly before calibration.
• Keep your scale in a clean environment. Dust, dirt, moisture, vibration, air currents
and vicinity to other electronic equipment (e.g. mobile phones) may cause an
adverse effect on the reliability and accuracy of the scale.
• Only operate the scale on a stable, horizontal and vibration free surface. Avoid
any draft.
• The scale has been calibrated in the factory. It may need user calibration from
time to time.
II. Operation
On initial use open the battery cover, remove wrapping foil from batteries and install
them in the in the compartment with electric contacts. Remove the protecting film on
the panel.
1. Press the ON/OFF button. Wait until [0.0] is displayed.
2. Gently place the load on the weighing platform.
3. Change the measuring unit with MODE if required.
III. Tare
1. The scale switched on, gently place a container on the weighing platform.
2. Press the TARE button and wait until [0.0] is displayed.
3. Fill in the net weight item(s) and directly read their weight.
IV. User calibration (only in gram unit)
1. Place the scale on a horizontal and stable surface.
2. Press ON/OFF and wait until [0.0] is displayed.
3. Press and hold MODE for approximately 3 seconds.
4. As soon as [CAL] is displayed release MODE.
5. Press MODE again. First [CAL] starts flashing followed by [500.0].
6. Carefully place a 500g calibration weight onto the centre of the platform.
7. [PASS] followed by [500.0] is displayed and the scale subsequently
returns to the normal weighing mode. End of calibration.
8. Remove the 500g calibration weight and [0.0] is displayed.
In case of inconsistency repeat the process.
Please check availability of optional weights with your dealer.
V. Elimination of problems and errors
Display
Elimination of errors
None
Remove isolation strip in battery compartment. Check polarity of
batteries. If still no display replace batteries.
[Lo_b]
Replace batteries. Pay attention to the polarity.
[UnSt]
Unstable: Move the scale to a horizontal and stable surface.
[OL_d]
Attention overload. Reduce the weight immediately.
[OUE2]
Calibrate the scale as described in point IV.
VI. Technical specifications
Capacity:
500g / 17.64oz / 7716gn / 2500ct
Resolution:
0.1g / 0.01oz / 1.5gn / 0.5ct
Units:
gram / ounce / grain / carat
User calibration:
Auto calibration with 500g weight
Tare range:
100% of weighing range
Auto shut-off:
After 120 seconds of no operation
ENGLISH