Calibration
button
Status window
Charging
Place the battery into the battery bay ensuring that the 5-way connector is fully seated.
The LEDs in the status window will provide status information and the charger will
automatically begin charging.
LED Indication:
The status of the battery is indicated by the LEDs visible in the status window:
Green flashing:
Green solid: Battery fully charged
Blue flashing:
Blue solid:
Red flashing: Battery fuel gauge in need of recalibration
Red solid:
Recharge Time
The ND2054 battery requires up to 3.5 hours to recharge a fully depleted battery.
Recalibration Time
Recalibration consists of a calibration charge, followed by a calibration discharge. Finally
the battery is given a regular charge. A calibration cycle will be faster if the battery is
fully charged to begin with. Recalibration time is governed by the battery voltage and
capacity. Larger batteries will take longer to recalibrate. Calibration is initiated each time
the button is pressed, so it is not recommended to press the recalibration button part
way through the recalibration cycle.
What are the SMBus and the SBS?
The Smart Battery System defines the parameters that are stored by an SBS-compliant
battery. These parameters include full battery status and fuel gauging information. The
®
LIT-171 CUDA
SURGICAL
(English)
Battery charging
Battery in calibration mode
Battery fuel gauge calibrated
Error
Battery Bay
DC
Connector
Rev. D
Date of Revision:3/28/13
Page 6 of 49