Troubleshooting - BrandTech Macro Mode D'emploi

Appareil sert au pipetage de liquides
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Trouble
Suction capacity
impaired
Pipette drips
Pipette not held
properly
Compressed suction
bellows take in air
automatically
Note:
Often visual control of the instrument is sufficient to discover a faulty
device which can easily be replaced by a spare part. The advantages of
do-it-yourself repair are obvious: It saves time and money.
Possible Cause
Filter dirty or valve
system damaged
Filter not properly posi-
tioned or valve system
damaged
Silicone adapter
damaged
Suction bellows or valve
system damaged

Troubleshooting

Action to be taken
Replace filter,
see pages 24-25
Insert filter properly or
replace valve system,
see page 26
Unscrew the adapter
support of the instru-
ment, replace silicone
adapter
Replace the looking ring
with the suction bellows
or the valve system
29