Specifications
Specifications subject to change without notice.
Application
Output temperature precision
Display resolution
Set point range
Flow rate range
Power requirements for single
phase (with grounded neutral) only
Installation category
Inlet pressure
Sample fittings
Weight
Wetted parts
Ambient operating temperature
Pollution Degree
Protection Class
Maximum Altitude
Enclosure Rating
Compliance Information
Storage Temperature
Humidity
1
Within proportional control range (see charts) and input temperature ±5 ºC, flow rate ±10%, 115 VAC, 60 Hz
±2 Hz and ambient temperature ± ºC. Instrument tuned at operating conditions (Autotune 3). Cal offset
adjusted to yield the set temperature at the point of use.
Sample heater operating range
Figure 1
shows in graph form the sample heater operating range in terms of sample output
temperature and minutes.
of sample output temperature and line AC voltage.
For water samples. Do not use the application for concentrated acids,
bases or any type of volatile sample.
±1.5 ºC (2.7 ºF) under reference conditions
1 ºC (1 ºF)
20 to 50 °C (68 to 122 °F) Refer to operating charts for more information.
50-300 mL/min; Sample Heater internal volume is 39 mL
120 Vac, 50/60 Hz, 11A
230 Vac, 50/60 Hz, 5.5A
II
689.5 kPa (100 psi) max. internal pressure. This rating is for the Sample
Heater and does not affect the Analyzer requirements.
Two ¼" OD Tubing Compression fittings
Net—4.98 kg (11 lb)
Shipping—6.35 kg (14 lb)
316 Stainless Steel
0 to 50 °C (32 to 122 °F)
2
I
2000 m (6560 ft)
NEMA 4X, IP 65
Listed for use in general locations to UL and CSA Standards by ETL. CE
certified.
-20 to 70 ºC (-4 to 158 ºF) with fluids drained.
Maximum relative humidity 80% for temperatures up to 31 ºC decreasing
linearly to 50% R.H. at 40 ºC.
Figure 2
shows in graph form the sample heater operating range in terms