LDO Sensor Replacement Kit
Each LDO sensor cap comes with a sensor-specific update module that calibrates
the meter for use with the sensor cap.
The kit includes:
Installing/Changing the Sensor Cap
Note: When installing a sensor cap for first-time use, begin with step 3.
1. To change the sensor cap, remove the old cap from the probe and discard.
2. Carefully remove the old seal by rolling it off with your fingers. To avoid damage
3. Position the new seal on the probe body with the narrow shoulder of the seal
4. Finger-tighten the new sensor cap onto the probe, just until the seal is
5. Visually inspect the probe seal to make sure the narrow shoulder of the seal is
6. Press SETUP . Use the UP and DOWN keys to highlight New Sensor in the
7. Press READ/OK.
Sensor cap.
Factory calibration update module.
Desiccant cartridges (packaged separately).
Avoid touching or soiling the exposed internal probe window.
If the internal probe window is soiled, rinse with dilute isopropyl alcohol or
deionized water and blot completely dry. Do not wipe the probe window or use
any abrasive cleaners.
Note: DO NOT use alcohol or other organic solvents to clean the black face of
the sensor cap. These solvents will destroy the sensor cap.
to the probe body, do not use sharp metal tools.
facing the probe tip. Slide the seal until it rests against the probe body threads.
See Figure 1. Do not touch the internal probe window.
compressed. Do not over-tighten. The cap will push the seal over the threads and
into position. See Figure 1 .
Note: Avoid handling the black face of the sensor cap.
inside the cap.
O
Setup Menu.