Guidance and manufacturer's declaration
electromagnetic emissions
The NEBULIZER MODEL NE1005 is intended for use in the electromagnetic environment
specified below. The customer or the user of the NEBULIZER MODEL NE1005 should assure
that it is used in such an environment.
Emissions
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Complies
flicker emissions
IEC 61000-3-3
The NEBULIZER MODEL NE1005 is intended for use in the electromagnetic environment
specified below. The customer or the user of the NEBULIZER MODEL NE1005 should assure
that it is used in such an environment.
Immunity test
Electromagnetic environment-
guidance
Electrostatic
The NEBULIZER MODEL NE1005
discharge (ESD)
uses RF energy only for its
IEC 61000-4-2
internal function. Therefore,
its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
Electrical fast
transient/burst
IEC 61000-4-4
T h e N E B U L I Z E R M O D E L
Surge
NE1005 is suitable for use in
IEC 61000-4-5
all establishments, including
domestic establishments
and those directly connected to
the public low-voltage
Voltage dips, short
power supply network that
interruptions and voltage
supplies buildings used for
variations
domestic purposes.
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level
Guidance and manufacturer's declaration
electromagnetic immunity
IEC 60601
Compliance level
test level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines and dc lines
supply lines and dc lines
±1 kV line(s) and neutral
±1 kV line(s) and neutral
<5% UT
<5% UT
(>95% dip in UT)
(>95% dip in UT)
for 0,5 cycle
for 0,5 cycle
40% UT
40% UT
(60% dip in UT)
(60% dip in UT)
for 5 cycles
for 5 cycles
70% UT
70% UT
(30% dip in UT)
(30% dip in UT)
for 25 cycles
for 25 cycles)
<5% UT
<5% UT
(>95% dip in UT)
(>95% dip in UT)
for 5s
for 5s
3 A/m
3A/m
E l e c t r o m a g n e t i c
e n v i r o n m e n t
-guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment. If
a dips or an interruption
of mains power occurs,
t h e c u r r e n t o f t h e
N E B U L I Z E R M O D E L
NE1005 may be dropped
off from normal level, it
may be necessary to use
uninterruptible power
supply or a battery.
P o w e r f r e q u e n c y
m a g n e t i c f i e l d s
should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.