14 Electrode Pads User Guide

Clean skin thoroughly prior to each application of electrode pads. Pads will not stick well if any
lotion, make-up, or dirt is left on your skin.
Make sure the Electronic Pulse Stimulator is off before applying the electrodes to the skin.
1.
Insert the pins of connecting wire into the wire connectors of each electrode pad (as
shown in image A)
2.
Remove the electrode pads from the protective liner. Lift at the edge of the electrode pad
and peel from the liner. Do not pull on the lead wires when removing the electrode pads
from the protective liner or skin. The electrodes may become damaged (show in image B).
3.
Apply firmly to the skin.
4.
When not in use, return the pads to the protective liner (as show in image C). Store the
electrode pads in the re-sealable bag in a cool place, out of direct sunlight.

15 Quality Assurance Tests

The test result of each step should meet the following pass criteria. Otherwise, it is considered
failed.
KEY FUNCTION TEST
When the functional keys are pressed step by step, the outputs of the device should be identical
to the following description:
1.
When the ON/OFF button at the off condition is pressed. The device is turned on, LCD
screen will show 7 modes. Minutes left, and A&B output channels.
2.
When the MODE key is pressed, the corresponding mode is selected on the LCD screen.
3.
Press the
0.
4.
The number in the top left corner display the remaining stimulation time.
LCD CHECK
The information displayed on the LCD screen should be identical to the following description:

16 Cleaning and Measurement

Use water or neutral detergent to clean the device first, and then use a dry cloth to wipe it
again. The electrode pads included with the device are reusable up to 80 uses. They should
button to decrease or increase the stimulation intensity. The default level is
Output Channels
Stimulation Timer
Level of Intensity
Battery indicator
current Stimulation Mode