mAINTENANcE, cARE, AND SERvIcE
Order all spare parts from your dealer or pharmacist, or contact Colpharma-Service (see foreword).
•
We recommend replacing the nebuliser, mouthpiece, and the face masks after 3 months use.
•
Check the filter continually for cleanliness and replace it if dirty, or after a maximum of 3 months use.
Spare filters are provided with the device.
•
To replace the filter, open the air filter compartment (4) on the top of the device marked «Filter» and exchan-
ge the filter (10).
Never unscrew the plastic lid.
mALFUNcTIONS AND AcTIONS TO TAKE
The device cannot be switched on
•
Ensure the power lead (2) is correctly plugged into the socket.
•
Ensure the ON/OFF switch (3) is in the position «I».
The nebuliser functions poorly or not at all
•
Ensure the air hose (6) is correctly connected at both ends.
•
Ensure the air hose is not squashed, bent, dirty or blocked. If necessary, replace with a new one.
•
Ensure the nebuliser (5) is fully assembled and the colored vaporiser head (11) is placed correctly.
•
Ensure the required medication has been added.
16