Electrical Specifications
Accuracy is ±(% reading + number of digits) at 23°C ± 5°C
< 80%RH.
(1) Voltage
Range
60.00mV
DC
600.0mV
DC
6.000V
DC
60.00V
DC
600.0V
DC
1000V
DC
60.0mV
AC
600.0mV
AC
6.000V
AC
60.00V
AC
600.0V
AC
1000V
AC
Input Protection : 1000V
AC Frequency Response : 50Hz ~ 1KHz
Input Impedance : 10MΩ, <100pF
AC Conversion Type :
AC conversions are ac-coupled, true rms responding,
calibrated to the sine wave input.
For non-sine wave add the following Crest Factor corrections:
For Crest Factor of 1.4 to 2.0, add 1.0% to accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to accuracy.
Resolution
0.01mV
0.1mV
0.001V
0.01V
0.1V
1V
0.01mV
0.1mV
0.001V
0.01V
0.1V
1V
DC
For 61-484
Accuracy
± (0.1% + 2d)
± (0.09% + 2d)
± (0.09% + 2d)
± (0.09% + 2d)
± (0.09% + 2d)
± (1.5% + 5d)
± (1.0% + 3d)
± (1.0% + 3d)
± (1.0% + 3d)
± (1.0% + 3d)
or 1000V
rms
AC
17
For 61-486
Accuracy
± (0.08% + 10d)
± (0.08% + 2d)
± (0.08% + 2d)
± (0.08% + 2d)
± (0.08% + 2d)
± (0.08% + 2d)
± (1.20% + 5d)
± (1.20% + 5d)
± (0.80% + 5d)
± (0.80% + 5d)
± (0.80% + 5d)
± (0.80% + 5d)