LaMotte 2020t/i Mode D'emploi page 15

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

ENTER
3.
Press
ENTER
4.
Press
Options.
ENTER
5.
Press
Calibration.
6.
Scroll to the desired calibration
option.
®
NOTE: Stablcal
standards below 50 NTU should not be used to calibrate the 2020t/i. The
diluent has a diff erent refractive index than traditional formazin standards and will aff ect
the results.
ENTER
7.
Press
The screen will display Storing... for
about 1 second and return to the
Options menu. Press
to a previous menu.
to select Options.
to select Turbidity
to select Turbidity
to save the selection.
EXIT
to return
Options Menu
Turbidity Options
Averaging
Set Clock
Set PWR Save
12:00:00
Turbidity Options
Turbidity Calibration
Turbidity Units
12:00:00
Turbidity Calibration
Nephelometric NTU
Ratiometric NTRU
Attenuation AU
12:00:00
Turbidity Calibration
Nephelometric NTU
Ratiometric NTRU
Attenuation AU
12:00:00
Options Menu
Turbidity Options
Averaging
Set Clock
Set PWR Save
12:00:00