450 Series Waterproof Handheld Meters
Dissolved Oxygen Operation Instructions
Models:
DO 450
DO 450
DO/Temperature
pH/mV/DO/Temperature
Getting Started/Connections
After installing (2) AA batteries and/or connecting the optional 110/220
VAC power supply, connect the desired sensors to the corresponding ports.
DO: 9-Pin RDO/ATC
ATC: 6-Pin ATC
(PD 450 only)
A/C Power
wer
Adapter
Output (USB or RS232 – PC
O
or Printer Communication)
o
The stand can be extended as shown above or used for wall-mounting.
Keypad Functions
Press once to power ON in the mode that was previously
used. Press again to turn backlight on for one minute or off.
Hold for 3 seconds to power OFF.
Toggle between measurement and calibration modes. In
SETUP mode, BACK serves to return to the previous menu
option or setting.
Confi rm calibration values in CAL mode.
Confi rm selections in SETUP mode.
Freeze or release the measured reading.
Customize instrument settings and preferences.
(See also Setup Programs)
Toggle between available measurement types.
Save measurement into memory. Increase value or scroll
up in SETUP or manual calibration.
Recall saved values from memory. Decrease value or scroll
down in SETUP or manual calibration.
Send output data to printer or computer.
Setup Programs
To access the settings below, press SETUP. Up/down arrows will display the
available options. Press ENTER to accept the desired setting, or BACK to
return to the previous option and/or exit.
Confi guration Options
• Ready indicator ON / OFF / or Automatic HOLD when stable
• Choose ºCelsius or ºFahrenheit
RDO Setup Options
• Choose units (% saturation, mg/L, or ppm)
• Choose number of calibration points 1 (no zero) or 2 (includes zero)
• Set compensation pressure in kPa (60.0 - 113.3), or mmHg, (450.0 -
850.0), salinity correction (0.0 - 45.0 ppt)
RDO Info
• Remaining Cap Life (# Days)
• Probe Serial Number (####)
• Year of Probe Manufacture (####)
• Date of Probe Manufacture (DD.MM)
Select Calibration Due Reminder
• Set number of days from 0-60
View Calibration Data
• Press ENTER to view each point that is calibrated.
View Electrode Data
• Press ENTER to view electrode data (Torr / % saturation).
System Settings
PD 450
PD 450
• Data Logging:
MANUAL upon key press only
TIMED interval. Choose (SEC / MIN / HOUR) interval.
• Automatic shut off after 10 minutes. Choose ON or OFF.
• Clock Settings:
Date: Choose USA (MM/DD/YYYY) or Euro (DD/MM/YYYY).
Time: Choose (24HR or 12HR). If 12HR, choose AM or PM.
• Set Printer Type:
CSV (Comma Separated Values) – best format for computer.
Printer (Text) – best format for printer.
Choose Manual (MAN) upon key press or TIMED interval.
If timed, choose (SEC / MIN / HOUR).
Reset
• NO. Exits from reset menu options without action.
BNC: pH, mV/ORP
• FACTORY RESET. Returns all settings except date/time and ATC
calibration to factory default values after ENTER is pressed then restarts
or Ion Selective
meter.
(PD 450 only)
• DATA RESET. Erases data stored in memory while retaining other
settings after ENTER is pressed.
• CALIBRATION RESET. Erases non-ATC calibration data while retaining
other settings after ENTER is pressed.
Dissolved Oxygen Calibration
Perform daily calibration for best results. Dissolved oxygen levels vary
with temperature, barometric pressure, and salinity, so calibration must be
performed with consideration of these factors. It is necessary to adjust the
barometric pressure and salinity values and calibrate the temperature prior
to performing any DO calibration or measurement. For probe operation,
refer to RDO Operation Instructions.
The 450 series can accept one or two calibration values; typically 100
% using saturated air or air-saturated water, and 0 % using zero oxygen
solution. If calibrating for 0 % oxygen, note that the meter will take several
minutes to reach 0 % saturation value and constant stirring is not required.
100 % Saturation Automatic Calibration
1. Rinse the probe well with clean water. From % saturation mode, press
CAL.
2. The primary display will show the calibration value, while the secondary
display shows the unadjusted value and the lower display shows the
measured temperature.
3. Place the probe in the optional calibration chamber or a saturated
oxygen environment (air or air saturated water) & wait for the reading to
stabilize. Press ENTER to accept.
0 % Saturation Automatic Calibration
1. Rinse the probe well with clean water. From % saturation mode, press
CAL.
2. The primary display will show the calibration value, while the secondary
display shows the unadjusted value and the lower display shows the
measured temperature.
3. Submerse probe in zero oxygen calibration solution and stir. Wait for the
reading to stabilize – this can often take more than 10 minutes. Wait for
the reading to stabilize. Press ENTER to accept.
mg/L or ppm Manual Calibration
1. Rinse the probe well with clean water. From mg/L or ppm mode, press
CAL.
2. The primary display will show the calibration value, while the secondary
display shows the unadjusted value and the lower display shows the
measured temperature.
3. Place the probe into a sample of known oxygen concentration (i.e.
determined by titration or another instrument). Wait for the reading to
stabilize. Press t/s to adjust the value and ENTER to accept.
Temperature Calibration/Manual ATC
1. Press CAL from any measurement, then press MODE.
2. Skip to step 3 for manual ATC, otherwise, dip the temperature sensor
into a solution with a known accurate temperature. The upper display
shows the active temperature while the lower display shows the factory
default temperature without adjustment.
3. Use up/down arrows to adjust the upper display. Press ENTER to accept
the calibration temperature. The maximum adjustable value is ±10 ºC
(or ±18 ºF) from factory default.
(for the parameter being measured)
(for the parameter being measured)
Error Messages
"ERR" will appear when an error condition exists or the incorrect
key is pressed. Common examples include:
• Pressing ENTER during calibration before the "READY" indicator appears.
Wait for the "READY" indicator before pressing ENTER.
• UR (Under Range) • OR (Over Range)
Intended Use, Maintenance & Precautions
These handheld meters use sensors to detect various parameters for water-
based measurements. For routine maintenance disconnect the power cord
or battery, then dust or wipe the display using a damp cloth. If necessary,
warm water or a mild water based detergent can be used. Immediately
remove any spilled substance from contact with the meter using the proper
cleaning procedure for the type of spill.
• Do not use this equipment in potentially explosive atmospheres.
• Refer to the electrode instructions for use, storage and cleaning.
• Ensure that no liquid enters the instrument.
• Do not use any aggressive cleaning chemicals (solvents or similar
agents).
• There are no user serviceable parts inside. Attempts to service internal
parts may void the warranty.
• Not intended for medical applications or patient use.
• WARNING: No modifi cation of this equipment is allowed.
Instrument Operating Conditions
Operating Ambient Temp.
5 to 45 ºC
Operating Relative Humidity
5 to 85 %, non-condensing
Storage Temp.
-20 to +60 ºC
Storage Relative Humidity
5 to 85 %, non-condensing
Pollution
Degree 2
Overvoltage
Category II
Weight
500 g
Size (L x W x H)
21.15 x 9.87 x 5.85 cm
Regulatory & Safety
CE, TUV 3-1, FCC Class A
Power Rating
DC Input: 9 VDC 1 A
2 x AA (LR6) 1.5 V batteries (replace
Battery Requirement
batteries when battery sign blinks)
Vibration
Shipping/handling per ISTA #1A
Shock
Drop test in packaging per ISTA #1A
Enclosure (Designed To Meet)
IP67 (using rubber covers)
Universal Power Adapter Operating Conditions
Operating Ambient Temp.
0 to 50 ºC
Operating Relative Humidity
0 to 90 %, non-condensing
Storage Temp.
-20 to +75 ºC
Storage Relative Humidity
0 to 90 %, non-condensing
Pollution
Degree 2
Overvoltage
Category II
I/P: 100 - 240 V, 50/60 Hz, 0.3A
Power Rating
O/P: 9 VDC 1 A
Oakton Instruments
625 East Bunker Court, Vernon Hills, IL, 60061, USA
Tel: 1-888-462-5866 • Fax: 1-847-247-2984
info@4oakton.com • www.4oakton.com
68X651903 09/15 Rev1

Publicité

loading

Sommaire des Matières pour Oakton 450 Serie

  • Page 1 RDO Setup Options into a solution with a known accurate temperature. The upper display shows the active temperature while the lower display shows the factory Oakton Instruments • Choose units (% saturation, mg/L, or ppm) default temperature without adjustment. • Choose number of calibration points 1 (no zero) or 2 (includes zero) 625 East Bunker Court, Vernon Hills, IL, 60061, USA 3.
  • Page 2 (p. ej. determinada mediante tritación u otro instrumento). • Indicador listo ON / OFF / o HOLD (Mantenimiento) automático si estable Espere a que se estabilice el valor medido. Presione t/s para ajustar Oakton Instruments • Elija ºCelsius o ºFahrenheit el valor y ENTER para aceptar.
  • Page 3 10 minutes. Attendre que la lecture se stabilise. Appuyer sur ENTER (Entrée) pour valider. Options de confi guration Oakton Instruments mg/L ou ppm Étalonnage manuel • Indicateur de préparation ON / OFF / ou automatiquement fi gé (HOLD) si 625 East Bunker Court, Vernon Hills, IL, 60061, USA stable 1.
  • Page 4 Lagerung Relative Luftfeuchtigkeit 5 bis 85 %, nicht kondensierend Bedienungsanleitung. Den gemessenen Wert einfrieren oder freigeben. Belastungs Grad 2 Die 450 Serie kann einen oder zwei Kalibrierungswerte akzeptieren; Überspannungs kategorie II normalerweise 100 % unter Verwendung von gesättigter Luft oder Instrumenteneinstellungen und Präferenzen anpassen.

Ce manuel est également adapté pour:

Do 450Pd 450