Manufacturer's declaration-electromagnetic emissions
The FocusTemp is intended for use in the electromagnetic environment (for
home healthcare and professional healthcare) specified below.
The customer or the user of the FocusTemp should assure that it is used in
such an environment.
Emission test
RF-emissies CISPR 11
RF-emissies CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
/flicker emissions IEC 61000-
3-3
Manufacturer's declaration-electromagnetic immunity
The FocusTemp is intended for use in the electromagnetic environment (for
home healthcare and professional healthcare) specified below.
The customer or the user of the FocusTemp should assure that it is used in
the environment specified below.
Immunity test
Electrostatic
discharge(ESD)
IEC 61000-4-2
Compliance
Group 1
Class B
Not applicable
Not applicable
IEC 60601
test level
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
EN-19
Electromagnetic
environment-guidance
(for home healthcare and
professional healthcare)
The FocusTemp uses RF
energy only for internal use.
Therefore, its RF emissions
are very low and are not likely
to cause any interference
from nearby electronic
equipment.
The FocusTemp is suitable
for use in all establishments,
including domestic
establishments and those
directly connected to the
public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Compliance
level
Contact:±8 kV
Air±2 kV,±4 kV,±8
kV,±15 kV
Electromagnetic
environment-
guidance (for
home healthcare
and professional
healthcare
environment)
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.