8.3. Compatibilité électromagnétique - Conseils et
déclaration du fabricant
Guidance and manufacturer's declaration - electromagnetic emissions
The LM-ProPower is intended for use in the electromagnetic environment specified below.
The customer or the user of the LM-ProPower should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The LM-ProPower is intended for use in the electromagnetic environment specified below.
The customer or the user of the LM-ProPower should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage varaiations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Group 1
Class B
Not applicable
Not applicable
IEC 60601
test level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for input/output
±1 kV for input/output
lines
lines
±1 kV differential
±1 kV differential
mode
mode
±2 kV common mode
±2 kV common mode
<5 % U
<5 % U
T
(>95 % dip in U
)
(>95 % dip in U
T
for 0,5 cycle
for 0,5 cycle
40 % U
40 % U
T
(60 % dip in U
)
(60 % dip in U
T
for 5 cycles
for 5 cycles
70 % U
70 % U
T
(30 % dip in U
)
(30 % dip in U
T
for 25 cycles
for 25 cycles
<5 % U
<5 % U
T
(>95 % dip in U
)
(>95 % dip in U
T
for 5 sec
for 5 sec
3 A/m
3 A/m
Garantie et déclaration de conformité
Electromagnetic environment - guidance
The LM-ProPower uses RF energy only for its internal function.
Therefore, it's RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The LM-ProPower is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Electromagnetic environment -
Compliance level
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
T
typical commercial or hospital
)
T
environment. If the user of the
LM-ProPower requires continued
operation during power mains i
T
interruption, it is recommended that the
)
T
LM-ProPower be powered from an
uninterruptible power supply or battery.
T
)
T
T
)
T
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
guidance
29