ELECTROmAGNETIC COmPATIBILITY (EmC)

EN
C
60601
This device must
be separated from
household waste
and recycled as
electronic waste
Expected service life is min. 400 hours.
E L E C T R O m A G N E T I C C O m P A T I B I L I T Y ( E m C )
Cryo Cuff IC has been tested and found to comply with the electromagnetic compatibility (EMC) limits
for medical devices to IEC 60601-1-2. These limits are designed to provide reasonable protection against
harmful interference in a typical medical installation .
CAUTION:
Medical electrical equipment requires special precautions regarding EMC and must be
installed and operated according to these instructions . It is possible that high levels of radiated or conducted
radio-frequency electromagnetic interference (EMI) from portable and mobile RF communications
equipment or other strong or nearby radio-frequency sources, could result in performance disruption
of the system . Evidence of disruption may include equipment ceasing to operate, or other incorrect
functioning. If this occurs, survey the site of disruption, and take the following actions to eliminate the
source(s).
Turn equipment in the vicinity off and on to isolate disruptive equipment .
Relocate or reorient interfering equipment.
Increase distance between interfering equipment and your system.
Manage use of frequencies close to the system frequencies.
Remove devices that are highly susceptible to EMI.
Lower power from internal sources within the facility control (such as paging systems).
Label devices susceptible to EMI.
Educate clinical staff to recognize potential EMI-related problems.
Eliminate or reduce EMI with technical solutions (such as shielding).
Restrict use of personal communicators (cell phones, computers) in areas with devices susceptible to
EMI .
Share relevant EMI information with others, particularly when evaluating new equipment purchases
which may generate EMI.
Purchase medical devices that comply with IEC 60601-1-2 EMC Standards (3 V/meter EMI immunity,
limit interference level to 0.0014 V/meter).