SPECIFICATIONS
Accuracy is specified for a period of one year after calibration and at 23°C ±5°C, with
relative humidity up to 75%.
Except where specified specially, accuracy is specified from 8% to 100% of range.
Accuracy specifications take the form of:
± ([% of Reading] +[number of Least Significant Digits)
AC Voltage
Range
Resolution
4.000V
1mV
40.00V
10mV
400.0V
100mV
600V
1V
Input Impedance: 10M
Frequency Range: 40Hz ~ 400Hz
Max. Allowable Input Voltage: 600V rms
Response: Average, calibrated in rms of sine wave
DC Voltage
Range
Resolution
400.0mV
0.1mV
4.000V
1mV
40.00V
10mV
400.0V
100mV
600V
1V
Input Impedance: range 400mV: >1000M
Max. Allowable Input Voltage: 600V dc
AC Current
Range
Resolution
400A
0.1A
Frequency Range: 50Hz ~ 60Hz
Max. Allowable Input Current: 400A
Response: Average, calibrated in rms of sine wave
Temperature Coefficient:
DC Current
Range
Resolution
400A
0.1A
Max. Allowable Input Current: 400A
Temperature Coefficient:
Accuracy
Overload Protection
+(1.2%+5)
600V rms
+(1.5%+5)
Accuracy
Overload Protection
+(0.8% + 5)
600V rms
+(1.0% + 5)
Accuracy
± (2.5% + 5)
Accuracy
± (2.5% + 5)