Cleaning the Charger
The charger cannot be disinfected. This procedure is for general cleaning as required, based on
physical appearance.
CAUTION: The charger is NOT waterproof. Do NOT immerse in water or any other
cleaning agent.
CAUTION: Dispose the charger according to the local regulations for battery disposal
(non-incineration).
To clean the charger:
1.
Wash your hands thoroughly.
2.
Use a damp cloth with mild cleaning solution, such as a dishwashing detergent, to clean any
dirt or foreign material from the outside of the charger. Never use organic solvents such as
paint thinner or acetone to clean the charger.
3.
Place the charger on a clean, dry cloth and air dry for 2-3 minutes.
Troubleshooting
Problem
I connected my device to
the charger and no lights
came on.
During charging, the
flashing green light on the
charger turns off and you
see a flashing red light on
the charger.
During charging, a mix of
quick and long flashing
red lights appear on the
charger.
Storage
Store the charger in a clean, dry location at room temperature.
Likely
Resolution
Cause(s)
Your
charger
Replace the battery in the charger.
battery has
no power.
Your
charger
battery
Replace the battery in the charger.
is low on
power.
Your
charger and
your device
Replace the battery in the charger and fully
batteries
charge your device.
are low on
power.