SciCan HYDRIM
Operator's Manual should be consulted before using this washer.
BASIC OPERATION
1
Power unit ON.
3
Close the door.
Refilling the Water Softener Salt Reservoir
Replacing the HIP
Recommended precautionary measures with regard to safe handling of HIP solution:
• Wear undamaged surgical gloves and eye protection.
• After handling the product, remove and dispose of the gloves and wash hands prior to removing eye protection.
TROUBLESHOOTING
Refer to Operator's Manual for a more comprehensive list. Call for service if suggested recommendations do not resolve the error message.
CF2 Chamber Filling
Water is not filling the chamber
CF15 Water Reservoir Overflow
Overflow switch triggered
CF21 Dosing System
Detergent is not dispensing
CF28
No Water Pressure
www.scican.com
MSDS available for download at www.scican.com
HYDRIM is a registered trademark and HIP is a trademark of SciCan Ltd. Manufactured by SciCan Ltd. 1440 Don Mills Road, Toronto, Ontario, Canada, M3B 3P9.
L110W G4
®
2
Load Instruments.
4
Select cycle and
press START.
1.
Unscrew lid.
2.
Fill to the top (max. 1 kg / 2.2 lbs.)
3.
Screw lid on tightly.
4.
Run a complete cycle with no instruments.
IMPORTANT!
If the salt container is being filled for the first time, it MUST be filled with
at least 1 litre of water before adding the salt.
Only tablet salt should be used in the softening unit. Fine grained salt
such as table salt should NOT be used as it could clog the unit.
The water softener salt is also available from SciCan.
Please contact your local dealer.
Cleaning Solution
Check that the water hoses are not kinked.
Check that water shut-off valves are opened.
Call for Service.
Restart the program.
Call for Service.
Tighten detergent cap.
Check for leaks.
Call for Service.
Check the water shut-off valves are open.
Check water hoses are not kinked.
Call for Service.
Quick Reference Guide
TOUCHSCREEN
P1
Rinse&Hold
P2
Regular
P3
Heavy Duty
Cleaning the Chamber
Pour 1.0L of vinegar or citric acid-based instrument neutralizer into
the chamber and follow these steps:
IMPORTANT! Do a Chamber Cleaning cycle every 25 cycles.
Failure to clean the chamber can promote corrosion.
Spare Parts and Accessories
CS-HIPL
01-112594S
01-108305S
WTI-SK
WTI-RK
FOR DETAILED INFORMATION PLEASE SEE PROVIDED OPERATOR'S MANUAL
1-800-667-7733
1-800-572-1211
Rinse and Hold Cycle
Prevents soil from drying on instruments.
Incomplete loads that will not be washed
within the hour.
NOT a cleaning cycle.
Regular Cycle
Moderately soiled loose instruments.
Heavy Duty Cycle
Heavily soiled instruments and cassettes.
Door
User/Technical
Lock
Menu
Network
USB
Connection
Water
Enter
Softener/
Detergent
Level/Door
Back
Status
P0 Cleaning
HIP Cleaning Solution (2 x 3.8L)
Water Softener Salt
Water Test Kits
Wash Test Indicator - Starter Kit
Wash Test Indicator - Refill Kit
Terminal sterilization of instruments MUST
follow processing in the HYDRIM L110W G4.