Les langues disponibles

CKW Serial Interface Option Kit
OPERATION
Sending Commands to the RS485 Interface
All commands to be sent to the specific RS485 must be preceded by the "@"
character and the two digit address set up in the COM menu. For example, to
send a print command to the RS485 that is setup as address 01, the user must
send the following sequence: "@01P".
Receiving Data from the RS485 Interface
Each response from the RS485 is preceded by the @ character and the two digit
address set up in the COM menu. For example, the same setup as above might
respond to the print command with the following reply: "@01 10.005 kg ".
Broadcast Commands
Using the address "00" with a command will act as a broadcast command, and
received by all systems on the RS485 network. Under normal circumstances there
is no response from a broadcast command. This is to insure that there is no
uncontrolled data being placed on the line.
If the address of a system is intentionally set to 00, then there may be a response
from this one system following a broadcast command. This option should be
used with care since the response will also contain the 00 broadcast address.

REPLACEMENT PARTS

Description
Serial Interface Printed Circuit Board
Long Option Cable Assembly
Short Option Cable Assembly

Service Information

For Service assistance in the United States and Canada, please call Ohaus
Corporation toll free at (800) 526-0659 from 8 AM to 5 PM Eastern Time. An
Ohaus Product Service Specialist will be available to help you.
EN-9
Part Number
71169391
71170778
71170779