Gima 28208 Mode D'emploi page 21

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 23
21
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by reflection from buildings, objects and people.
a) The field intensity for fixed transmitters such as the base stations for radiotelephones (mobile and cordless) and terres-
trial mobile radio, amateur radio devices, radio AM and FM transmitters and TV transmitters can not be theoretically and
accurately foreseen. To establish an electro-magnetic environment generated by fixed RF transmitters, an electro-magnet-
ic study of the site should be considered. If the field intensity measured in the place where the device will be used surpass-
es the above mentioned applicable level of conformity, the normal functioning of the device should be monitored. If abnor-
mal performance arises, additional measures such as changing the device's direction or positioning may be necessary.
b) The field intensity on an interval frequency of 150 kHz to 80 MHz should be less than 10 V/m.
Recommended separation distance between portable and mobile radio-communication devices and the monitor
The SUPER TOBI SUCTION UNIT surgical aspirator is intended to operate in an electro-magnetic environment where
RF irradiated interferences are under control. The client or operator of the SUPER TOBI SUCTION UNIT device can help
prevent electro-magnetic interference by keeping a minimum distance between the portable and mobile RF communi-
cation devices (transmitters) and the SUPER TOBI SUCTION UNIT device, as recommended below, in relation to the
radio-communication maximum output power.
Maximum nominal
output power of the
Transmitter W
0,01
0,1
1
10
100
For transmitters with a maximum nominal output power not shown above, the recommended separation distance in metres
(m) can be calculated using the equation applicable to the transmitter frequency, where P is the maximum nominal output
power of the transmitter in Watt (W) depending on the transmitter's manufacturer.
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is influenced by the
absorption and by the reflection from buildings, objects and people
Separation distance from the frequency transmitter (m)
150KHz to 80MHz
3,5
√P
d=
V
1
0,12
0,38
1,2
3,8
12
ENGLISH
Where P is the maximum nominal output voltage of the
transmitter in Watt (W) depending on the manufactur-
er of the transmitter and the recommended separation
distance in metres (m). The intensity of the field from
the fixed RF transmitters, as determined by an elec-
tro-magnetic study of the site
level of conformity of each frequency interval
It is possible to check for interference in
proximity to devices identified by the fol-
lowing symbol:
80MHz to 800MHz
12
√P
d=
E
1
0,12
0,38
1,2
3,8
12
, could be lower than the
a)
.
b)
800MHz to 2,7GHz
23
√P
d=
E
1
0,23
0,73
2,3
7,3
23

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

28224

Table des Matières