Cover and cover seal
1
1
Remove the cover by gently flexing one cover arm away and up from the docking
station until the cover detaches completely.
2
Peel the cover seal away from the cover.
3
Inspect the cover and cover seal weekly for wear and deterioration. Replace any
damaged parts. When required, wash the cover and cover seal in warm water, using
a mild detergent. Rinse them thoroughly with clean water and allow it to dry away
from direct sunlight.

Docking station

Periodically wipe the outside of the docking station (including the connector seal) with
a damp cloth.
CAUTION
!
Do not use any of the following to clean any part of the H4i as these solutions
may cause damage and reduce the life of the product:
bleach, chlorine-, alcohol- or ammonia-based solutions
moisturising, antibacterial or glycerine based soaps
water softening and unapproved descaling agents.
Heater plate
Do not allow water to collect on the heater plate. Wipe it with a clean cloth if
necessary.

Reassembling the H4i cleanable water chamber

1
1
Refit the tub plate seal into its groove on the cleanable water chamber.
2
Reinsert the tub plate by slotting its feet into the catches.
3
Reassemble the catch by lifting up and over the water chamber and then pushing
down to lock into place.
Note: Only the H4i Cleanable Water Chamber can be disassembled/reassembled.
2
2
3
Cleaning and maintenance