• Do not operate in the presence of flammable anesthetic mixtures, oxygen, or nitrous oxide.
• Do not operate the unit without the plastic grate on the bottom.
• Do not touch the bottom or sides of the metal tank, or a burn may result.
• Do not heat paraffin on a stove, hotplate, or in a microwave.
• Do use this appliance on a level, secure surface.
• Do not use any additives in the paraffin except as allowed by the manufacturer.
• Do not modify this equipment without authorization from the manufacturer.
• Care must be taken when operating this equipment around other equipment. Potential electromagnetic
or other interference could occur to this or the other equipment. Minimize this interference by not
using other equipment in conjunction with it.
Cautions
• Do not remelt used paraffin, reintroduced particles or chemicals may damage the tank.
• Never use sharp objects to remove hardened paraffin from the unit.
• Do not leave an empty unit plugged in.
• Do not pour melted paraffin from the tank.
NOTE: The emissions characteristics of this equipment make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally
required) this equipment might not offer adequate protection to radio-frequency communication services.
The user might need to take mitigation measures, such as relocating or re-orienting the equipment.

Technical Information

• Power, Model TB6: 100–125V, 0.87–1.3A, 50–60Hz
• Power, Model TB7: 220–240V, 0.55–0.65A, 50–60Hz
• Heater: TB6: 87–162 watts, TB7: 121-156 watts
• Operating temperature: 126º–134º F. (52.2º–56.7º C.)
• 14.3" L x 8.2" W x 7.5" H (36.32 cm L x 20.83 cm W x 19.1 cm H)
• Capacity: 9 lbs. (4.08 kg.)
• Melt time: 6-8 hours
• Duty cycle: continuous operation
• Degree of protection against moisture ingress: IPX0
• 220V unit must be used with a hospital grade cord of at least the maximum ratings: 10 A, 220 VAC.
• Meets IEC/EN 60601-1-2:2014 electromagnetic compatibility interference safety standard.

Environmental Ratings

Temperature:
Relative Humidity:
Atmospheric Pressure:
No.
Test/Standard
CISPR 11, Radiated Emissions
1
IEC 61000-4-2
2
IEC 61000-4-3 Radiated RF
3
Immunity
IEC 61000-4-3 Proximity fields from
4
RF wireless equipment
IEC 61000-4-4
5
IEC 61000-4-5
6
IEC 61000-4-6
7
IEC 61000-4-8
8
IEC 61000-4-11 Voltage Dips
9
IEC 61000-4-11 Voltage Interruptions
10
Operating
65° to 85° F (18.3° to 29.4° C)
10% to 95%
70 to 160 kPa
Port (enclosure,
AC, DC, I/O)
Enclosure
Enclosure
Enclosure
Enclosure
AC
AC
AC
Enclosure
AC
AC
• 2 / 12 •
NOTE: "Hospital Grade" grounding
reliability can only be achieved when
the equipment is connected to
an equivalent receptacle marked
"Hospital Only" or "Hospital Grade"
NOTE: Current and wattage
must be measured at operating
temperature.
Transport and Storage
-40° to 130° F (-40° to 54.4° C)
10% to 95%
70 to 160 kPa
Emissions Class and Group /
Immunity Test Level
Class A; Group 1
±8kV contact, ±2kV, ±4kV, ±8kV, ±15kV air
3V/m, 80-2700MHz, 80% 1kHz AM
Section 8.10 of the IEC 60601-1-2 standard
2kV, 100kHz repetition rate
Line-to-Line: ±0.5 kV, ±1 kV
Line-to-Ground: ±0.5 kV, ±1 kV, ±2 kV
3V, 0.15-80MHz, 80% 1kHz AM, 6V in ISM Band
within 0.15-80MHz, 80% 1kHz AM
30A/m, 50Hz and 60Hz
0% (100% reduction), 0.5 cycle; 0% (100% reduction),
1 cycle; 70% (30% reduction) UT, 0.5 sec
0% (100% reduction), 5 sec