Important; Other Operatinng Controls - Aqualytic AL654 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Trouble shooting
If the verification of the temperature accuracy is not successful the following should
be taken under consideration:
-
The environmental temperature during the verification has to be between +10 and
+32°C (climate class).
-
A certain circulation of air is necessary for the compressor to work properly.
-
The incubator should not be opened to often and / or for longer period of time.
-
Check the function of the two fans by visual inspection.
-
Check the cooling device by setting 10°C and wait for a significant change of the
temperature in the sample towards 10°C.
-
Check the heating device by setting 40°C and wait for a significant change of the
temperature in the sample towards 40°C.
-
Depending on the sample temperature it can take up to 7 hours for temperature stability
at the set-value.

Important:

The air temperature in the cabinet may fluctuate by several degrees centigrade.
Nevertheless the actual temperature of the sample fluctuates only by a maximum of
± 0.5°C.
4.5
Other operation controls
On the front side of the temperature control unit there are covered shock-
proof sockets for the connection of BOD's appliances or stirring apparates.
The total load on the sockets must not exceed 5 amps.
The thermostatically controlled oven and laboratory sockets are protected by a fuse
(6.3 A T) which is accessible from the front panel of the unit.
The control unit is connected by means of a 4-pole plug on the bottom of the incubator.
Important:
Please make sure that no additional heatsource (like a lamp) is installed!
22

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Al654gAl655Al655gAl656Al656gAl658 ... Afficher tout

Table des Matières