3.4. Installation of the appliance

When chosing a location for the appliance the following points should be taken into
account:
The floor of the working area should be flat and level. Any uneveness can be compen-
sated with the aid of the two adjusting feet (wrench size 22 mm).
The rear side of the appliance may be placed directly against the wall. The convection
grids on the rear side of the appliance are not to be covered (e.g. table tops etc).
To comply with European standard EN 378 the room in which the unit is installed must
have a volume of 1 m
there is a leak in the cooling agent circuit, no explosive mix of gas and air can develop
in the room. The quantity of cooling agent used in the unit is stated on the data plate
inside the unit.
The carrying capacity of the retractable grids are about 45 kg.
Overall dimensions of the appliances in mm:
AL654
AL654G
AL655
AL655G
AL656
AL656G
AL658
AL658G
• no direct exposure to sunlight
• no proximity to any form of heating
• well ventilated rooms
for every 8 g of R 600a cooling agent used in the unit. Thus, if
3
Height
850
885
1216
1216
1590
1590
1516
1516
Width
Depth
600
600
600
600
600
600
600
600
600
600
600
600
752
710
752
710
17