RECOMMENDED SEPARATION DISTANCE between portable and mobile RF
communications equipment and the Mesh–nebulizer (membrane) AGU N7.
The Mesh–nebulizer (membrane) AGU N7 is intended for use in an electromagnetic envi-
ronment in which radiated RF disturbances are controlled. The customer or the user of the
Mesh–nebulizer (membrane) AGU N7 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equip-
ment (transmitters) and the AGU N7 as recommended below, according to the maximum
output power of the communications equipment.
Rated maxi-
Separation distance according to frequency of transmitter, m
mum output
power
150 kHz to 80 MHz
of transmit-
d = 1,2 √P
ter, W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommend-
ed separation distance d in metres (m) can be estimated using the equation applica-
ble to the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
66
80 MHz to 800 MHz
d = 1,2 √P
0,12
0,38
1,2
3,8
12
800 MHz to 2,5 GHz
d = 2,3 √P
0,12
0,38
1,2
3,8
12
0,23
0,73
2,3
7,3
23