Annexe 4 Tableau de réglage utilisateur DO
* Référence dans le manuel IM 12A01A02-01E
Paramètre
Sensor
Sensor type
setup
Measure-
Configure
Unit
ment setup
sensor
Sensor sensitivity
Polarization Voltage
Tem-
Temp. element
perature
Unit
settings
Temp.
Compensation
compensa-
tion
Salinity
Compensation
compensa-
Input Salinity
tion
Pressure
Input Pressure Level (Process)
comp.
(Measure)
Calibration
Limits and
settings
timing
zero/slope
Pressure
Comp. (Cal.)
Zero Calibration
Sensor
Progress time:
diag.
settings
Heat cycle:
Define heat
cycle
Annexe 4 Tableau de réglage utilisateur DO
Valeurs par
Galvanic
mg/L
Gal: 0.45
µA/ppm(50
µm)
Pol: 7.5
nA/ppm
0.675 V
Pt1000
ºC
Automatic
Manual
Manual temp.
25.0 ºC
(77.0 ºF)
Disabled
0.0 g/kg
101.3 kPa
Zero
Low
Gal: -0.50 µA
Pol: -5.0 nA
High
Gal: 0.50 µA
Pol: 5.0 nA
Slope
Low
40.00 %
High
200.0 %
Stabilization time
60 s
Step
Zero
0.10mg/L, ppm
Range
1.0 ppb, %SAT
Span
0.10mg/L, ppm
1.0 ppb, %SAT
Cal.Interval
250 days
Zero Current
Gal: 0.00 µA
Pol: 0.0 nA
Slope
100.0 %
Input Pressure Level (Cal.)
101.3 kPa
Disabled
Disable
BAD limit
2000 days
Disable
BAD limit
500 cycles
Heat cycle tempe
50 ºC
Hear cycle time:
10.0 min.
Valeurs applicables
défaut
Galvanic, Polaro-
graphic
mg/L, ppm, ppb,
%SAT
Gal: 0.45 µA/ppm(50
µm), 0.90 µA/ppm(25
µm), Others (0.10 à
20.00 µA/ppm)
Pol: 1.0 à 200.0 nA/
ppm
0.000 à 1.000 V
Pt1000, NTC22k
ºC, ºF
Manual, Automatic
-20 à 150 ºC
Disabled, Enabled
0.0 à 100.0 g/kg
0.0 à 999.9 kPa
Gal: -20.00 à 0 µA
Pol: -200.0 à 0 nA
Gal 0 à 20.00 µA
Pol: 0 à 200.0 nA
0.1 à 99.9 %
100.0 à 600.0 %
5 à 600 s
ppm, mg/L:
0.00 à 50.00
ppb: 0 à 2000
%SAT: 0.0 à 600.0
1 à 250 jours
Gal: -20.00 à 20.00 µA
Pol: -200.0 à 200.0 nA
0.1 à 600.0 %
0.0 à 999.9 kPa
Disabled, Enabled
Disable, Enable
1 à 10000 jours
Disable, Enable
10 à 1000 cycles
1 à 250 ºC
0.1 à 60.0 min.
App.4-1
Réglages
*Réf.
utilisateur
16.1
16.2.1
16.2.2
16.2.3
16.2.4
16.2.5
16.2.6
16.2.7
DO
IM 12A01A02-12F-E