CONTROL PANEL & OPERATIONS
1.
– Working Time Display. It counts down after work is started.
10:00
2.
– Multistage Thermometer. Solid green light indicates: water temperature is 0-30% of
60 ºC/140 ºF. Solid orange light indicates: water temperature is 30-60% of 60 ºC/140 ºF. Solid red light
indicates: water temperature is 60-90% of 60 ºC/140 ºF. Flashing red light indicates: water temperature
is close to 60 ºC/140 ºF. All lights indicates: water temperature is at 60 ºC/140 ºF (90-100%).
3. Warning Light. Normally it could be illuminated in the following two scenarios:
a.
Overheat protection is triggered for one or two transducers. If it is only one transducer, the
other one will continue to work. Once the transducer is cooled down, the warning light will be
dimmed.
b.
If the light is illuminated as soon as the unit is turned on, it indicates malfunctioning at one or
two signal cables for transducer PCBs. The unit can be operated normally.
4.
– Normal Cleaning Status.
5.
– Degas Status. Illuminated, the degas function is selected. Flashing degassing is underway.
6. Timer Quick Set Key. Touch
increase it by 5 minutes.
7.
– Time Decreasing Key. Touch
8. Start/Stop Key. After powering on, the LED display shows 10:00, which is the initial
default time (as shown), and the unit is ready with the normal working status. Touch
Start/Stop key twice and the cleaning starts. The cleaning stops when the timer counts
down to 00:00. If the unit needs to be stopped before the timer runs out, touch Start/
Stop key twice.
To prevent accidentally turning on or off the unit, the Start/Stop key needs to be touched within 6 seconds
to turn it on or off.
9. Degas Key. Touch Degas key,
key twice to start degassing.
function. Touch Degas key before it ends will stop degassing.
Degas: When cleaning heavily soiled or greasy items or silver, copper or brass items, chemical solutions
need to be added into the water. The solution may form many small droplets and take a long time to
dissolve in the water. Newly added water may generate many air bubbles on the tank walls. These will
reduce the cleaning effect in the beginning phase of ultrasonic cleaning. Turning on the degas function will
dissipate the droplets and the air bubbles, usually in 90 seconds, and improve cleaning efficiency.
4
Illuminated, normal cleaning is underway.
Time+, it displays
key each time reduces the timer by 1 minute.
flashes, indicating the unit is running under Normal Cleaning mode.
illuminates, indicating Degas function is selected. Touch Start/Stop
will flash for 90 seconds, then the unit will return to the normal cleaning
or other quick timer settings. Each touch
10:00