F u n c t i o n Te s t

Function Test

This test should help verify, at certain intervals,
the functions of HM8012 without any great expen-
diture in measurement instruments. To achieve
thermal balance, the module and the basic instru-
ment, in its case, must be energized for at least
60 minutes before the test begins.

Measurement equipment used

Calibrator AC/DC for instance Fluke 5101B / Fluke
5700A / Rotek 600
Resistors of 5 kΩ, 50 kΩ, 500 kΩ 0.01% for instance
model S102 J by Vishay
Resistors 500 kΩ, 5 MΩ 0.02%, for instance models
CNS020 by Vishay.
Test procedure
If one of the indicated calibrators is available or
if precision calibrators are appropriate, all the
HM8012 measurements ranges can be checked
using the following tables which indicate the limit
values. Recalibration, however, should only be
performed if the appropriate precision calibrator
is available.
Before any change of ranges, ensure that the
signal at HM8012 does not represent an unaccep-
table load of the object under examination. For the
link between the calibrator and HM8012, shielded
cables must be used to prevent any unwanted
influence caused by the measurement signal.
a) DC voltage ranges
No. Range
Reference
(+23 °C)
1
500 mV
250 mV
2
5 V
2.5 V
3
50 V
25. V
4
500 V
250 V
5
600 V
550.00 V
32
Subject to change without notice
Display limits
249.85 - 250.15
2.4986 - 2.5014
24.985 - 25.015
249.86 - 250.14
549.7 - 550.3
b) AC voltage ranges
No. Range
Reference
(+23 °C)
1
500 mV
250 mV
2
5 V
2.5 V
3
50 V
25 V
4
500 V
250 V
5
600 V
550 V
(1) = 40 Hz to 5 kHz
(2) = 20 Hz to 20 kHz
(3) = 40 Hz to 1 kHz
(4) = 20 Hz to 1 kHz.
c) DC current ranges
No. Range
Reference
(+23 °C)
1
500 µA
250.00 µA
2
5 mA
2.5000 mA
3
50 mA
25.000 mA
4
500 mA
250.00 mA
5
10 A
1.800 A
d) AC current ranges (f = 400 Hz)
No. Range
Reference
(+23 °C)
1
500 µA
250.00 µA
2
5 mA
2.5000 mA
3
50 mA
25.000 mA
4
500 mA
250.00 mA
5
10 A
1.800 A
e) Resistor ranges
No. Range
Reference
(+23 °C)
1
500 Ω
200.00 Ω
2
5 kΩ
2.0000 kΩ
3
50 kΩ
20.000 kΩ
4
500 kΩ
200.00 kΩ
5
5 MΩ
2.0000 MΩ
6
50 MΩ
20.000 MΩ
Display limits
(1) 248.65 - 251.35
(2) 247.15 - 252.85
(1) 2.4865 - 2.5135
(2) 2,4715 - 2.5285
(1) 24.865 - 25.135
(2) 24.715 - 25.285
(3) 248.65 - 251.35
(4) 247.15 - 252.85
(3) 547.3 - 552.6
(4) 544.0 - 555.9
Display limits
249.48 - 250.52
2.4948 - 2.5052
24.948 - 25.052
249.48 - 250.52
1.794 - 1.806
Display limits
247.9 - 252.1
2.479 - 2.521
24.79 - 25.21
247.9 - 252.1
1.775 - 1.825
Display limits
199.83 - 200.17
1.9989 - 2.0011
19.989 - 20.011
199.89 - 200.11
1.9939 - 2.0061
19.393 - 20.061