c Perform the same operation again to
set up the second calibration temperature
such as 100 ;
Note: when setting up the temperature
points, an order of low temperature point
first and high temperature after is
recommended. If on the case high
temperature point is set first and low
temperature is set after, then when
performing the following actual calibration,
please follow the order of low temperature
first and high temperature after.
d) Press both "
same time to start the calibration. The
display will show 40.0 and the temperature
of the instrument will automatically
increase to 40°C. Once the temperature is
kept constant, timing will start. 30 minutes
later, press "
display will show the blink of "TEMP." At
this time, the display will show the actual
temperature reading of the thermometer.
For example, if the thermometer reads
40.5°C, you can adjust the temperature
display to 40.5°C by pressing the
temperature set up button. Next press "
"for one time, and the display will show
100.0°C. Following the same instruction,
when the temperature of the instrument
increases to 100°C, wait for 30 minutes,
" and "
" at the
" for one time, and the