EMC Tables

Electromagnetic emission
Emission test
RF emissions
Cispr 11
RF emissions
Cispr 11

Electromagnetic immunity

The device is inteded for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that is used in suche environment
Immunity test
Electrostatic discharge
(ESD)
EN 61000-4-2
Mains power
electromagnetic field
EN 61000-4-8
IACER S.r.l.
Compliance
Group 1
Class B
Test level EN
Compliance level
60601-1-2
 6kV contact
 6kV contact
 8kV air
 8kV air
3 A/m
3 A/m
32 di 36
Electromagnetic environment –
guidance
The device uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment
The device is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes.
Electromagnetic environment
– guidance
Floors sholud be wood, concrete
or ceramic tile. If floor are
covered with syntethic material,
the relative humidity should be
at least 30%
Mains power quality should be
at that of a typical commercial or
hospital environment
MNPG164 Rev. 02 02/07/15