the intake tube.
4.
Hold the dispense nozzle over the aperture of the reservoir and apply gentle piston strokes in order to return any syringe contents
into the reservoir.
5.
Empty the instrument completely and flush thoroughly with distilled water.
6.
If the piston barrel is still not completely clean, you need to dis-assemble the dispenser. Refer Dis-assembling procedure given
below.

Dis-assembling the Dispenser for Cleaning and Servicing

Procedure to dis-assemble the PISTON
1- Pull the cap outwards to expose the Calibration Nut.
2- Unscrew the Calibration Nut with the help of calibration tool to dis-assemble the Piston and shaft out of the main housing.
3- Unscrew the piston from the shaft
Procedure to dis-assemble the BARREL
1- Unscrew the Barrel Cap using the special tool provided with the instrument.