5. The Lower Menu
The functions: Unit 1, Unit 2, CAL 1 and CAL 2 can be selected in
the lower menu. Select with
confirmed by means of
. The menu can be cancelled by
not pressing for 20 seconds.
Unit 1: With Unit1 you select the unit of temperature. You have a
choice of °C and °F. You can select with
Unit 2: With Unit2 you select the unit of the rel./abs. humidity or the
dew-point temperature. You have a choice of g/m
You can select with
and
B - 05
. The selected function flashes and is
and
; confirm by
, %r.h. dp°C, dp°F.
3
and confirm with
.
Operating manual – English
or by
.
Single-point calibrations of
temperature and rel. humidity
CAL1: With CAL1 (single-point calibration)
you set the offset for Sensor 1 (temperature).
The offset is shown in the lower part of
the display. Maximum offset is ± 10 °C or
± 10 °F.
You can select with
and
with
.
The original works settings can be
achieved by setting the offset to 0.0.
and confirm