Télécharger Imprimer la page

Labconco CentriVap Guide D'installation Rapide page 2

Publicité

Les langues disponibles
  • FR

Les langues disponibles

CentriVap
Note: This Quick Guide to Installation may not provide all the information for
your installation. Refer to the User's Manual for details and precautions. Do not
discard packaging until you have identified all parts that were shipped and the
products operate properly.
Preparations
Step 1. Proper Electrical Supply
230V, 50 Hz models
o 15-Amp dedicated single phase circuit. Additional
appliances on the same circuit may cause the current to
exceed the rating of the circuit breaker or fuse.
o Frequency must agree with the serial tag rating.
Improper frequency will damage the compressor.
Step 2. Vacuum Pump Requirements
 For aqueous samples
o 90 liters per minute free flow rate (5,2 m
o 0,002 mBar ultimate vacuum
 For volatile solvent samples
o 42 liters per minute (8,6 m
o 200 mBar ultimate vacuum is adequate
o PTFE
 Plug Requirements on vacuum pump power cord
o 230V models - reverse IEC plug
 Hose barb connection to fit 13mm (0.5 inch) inside diameter
vacuum hose
Step 3. Vacuum Trap Requirements
 When using a mechanical pump that is not corrosion resistant,
the CentriVap Cold Trap or equivalent trap must be used in the
vacuum line to prevent damage from solvent vapors. When
evaporating acids, it is advisable to include a soda lime acid
secondary trap in the vacuum line. This adds additional
protection for the pump
®
Centrifugal Concentrator Systems
Quick Guide to Installation
3
per hour)
Page 1 of 4
3
per hour)
Reverse IEC plug
UK plug
EU plug

Publicité

loading

Produits Connexes pour Labconco CentriVap