10 .. R esidual. water. on. the. surface. of. the. sample. .
(e g .dew).may.have.a.direct.impact.on.the.results,.
so.you.should.avoid.such.situations.and.ventilate.
the.sample.before.taking.measurements
11 .. W ith.high.humidity.levels.(exceeding.25%).you.
should. avoid. pouring. very. cold. grain. into. the.
chamber,.as.water.vapour.liquefies.on.the.surface.
of. the. grain . In. such. situations. mix. thoroughly.
the. portion. of. grain. from. which. the. sample. is.
taken. and. wait. several. minutes. until. the. grain.
temperature. increases . The. first. measurement.
result. should. therefore. be. ignored. and. the.
final. result. should. be. the. average. of. the. three.
subsequent.consecutive.measurements
12