Safety instructions
This instrument is designed for use in Explosion Hazard
Zone 2 areas. Do not use in Explosion Hazard Zone 1 areas.
If you are working in explosion hazard environments,
do not use plastic beakers. Plastic beakers can create sta-
tic electrical charges and can ignite
explosive substances. The power
supply plug must be connected up
outside the explosion hazard zone.
The housing must be earthed. To
earth the instrument, use a ear-
thing cable with a commercial 6.3
quick-connect terminal and plug
this into the socket on the rear of
the instrument (see figure).
Damage to the protective membra-
ne sheet will reduce the IP and explosion hazard zone
rating of the instrument.
Hazard warning – Magnetism. The instrument creates a mag-
netic field. Items which are sensitive to magnetism (e.g. da-
ta storage devices, pacemakers...). must be kept away from
6
the instrument.
Correct use
Page
The IKA
mixing up to 800 ml of thinly fluid media with magnetic
6
bars up to 30 mm long. Stirring speed can be adjusted in
6
the range from 15 to 1500 rpm.
6
6
Unpacking
7
7
Please unpack the device carefully and check it for dama-
8
ge. It is important that any transport damage is detected
when the device is unpacked. If necessary, an inventory
of the damage should be made immediately (postal ser-
vice, railway, haulage company).
As supplied, the package contains:
The IKA
tic bar, two beakers and an instruction manual.
Commissioning
Plug the instrument into a power supply socket. (For the
power supply specifications, see the section "Technical
Data").
Place the magnet bar and the medium to be mixed in a
container which is transparent to magnetic fields. Set the
stirrer speed to minimum. To do this, switch on the
instrument by pressing the
the
and plus buttons stops flashing and lights permanently.
You can now place the container on the mark in the cen-
tre of the set-up surface.
®
lab disc is a magnetic stirrer designed for
®
lab disc, a plug-in power supply unit, a magne-
button and then press
button until the flashing LED between the minus