1.
Do not drop this monitor nor subject it to strong impacts.
2.
Avoid high temperatures and direct sunlight. Do not immerse the monitor in
water as this will result in damage to the monitor.
3. If this monitor was stored at a temperature near the freezing point, allow it to
come to room temperature before use.
4.
Do not attempt to disassemble this monitor.
5. If you do not use the monitor for a long time, please remove the batteries.
6. It is recommended the performance should be checked every 2 years or after
repair. Please contact the service centre.
7. Clean the monitor with a dry, soft cloth or a soft cloth squeezed well after mois-
tened with water, diluted disinfectant alcohol, or diluted soapsuds.
8. No component in the monitor can be serviced by the user. The circuit diagrams,
component part lists, descriptions, calibration instructions, or other information
which will assist the user's appropriately qualified technical personnel to repair
those parts of equipment which are designated repairable can be supplied.
9. The monitor can maintain the safety and performance characteristics for a mini-
mum of 10,000 measurements or three years, and the cuff integrity is maintained
after 1,000 open-close cycles of the closure.
10. It is recommended that the cuff should be disinfected 2 times every week if
needed (for example, in a hospital or in a clinic). Wipe the inner side (the side
that contacts skin) of the cuff with a soft cloth moistened with ethyl alcohol
(75 - 90 %) and squeezed, then dry the cuff by airing.
Explanation of Symbols on Appliance
Symbol for "THE OPERATION GUIDE MUST BE READ"
(Symbol = white with blue background)
Symbol for "WARNING"
Symbol for "TYPE BF APPLIED PARTS" (the cuff is a type BF applied part)
Symbol for "ENVIRONMENT PROTECTION" - Waste electrical products
should not be disposed of with household waste. Please recycle where fa-
cilities exist. Check with your local Authority or retailer for recycling advice.
Symbol for "MANUFACTURER"
116
PC-BMG3019_IM
Maintenance
18.05.18