EN
1.2. Installation
Independent leak test station:
1.
Connect TGF10 to Leak Detector
Sensistor ISH2000.
2.
Connect compressed air to compressed
air port.
3.
Connect Tracer gas port to gas bottle.
4. Connect a large hose (∅ 12 mm. or more) to the exhaust port. Remote end of hose should be placed in
a ventilation stud carrying out of the building.
NOTE: Maximum length for a 12 mm hose is 5m. Use bigger diameter if longer hose is needed. Too
high resistance in the exhaust hose will severely reduce capacity of venturi pump.
5.
Connect Test port to test object or fixture.
Adding leak detection equipment to an existing pressure decay station:
1 - 4 As above.
5
Connect test port to exhaust port of pressure decay tester.
NOTE. Test port of TGF10 is open to exhaust in standby mode. This means that the pressure decay
instrument can operate normally with the TGF10 instrument connected.
4
INFICON – User's manual TGF10