Cleaning and sanitising the medicinal product reservoir
Open the vessel and discard any leftover medicinal product.
Pour a small amount of hot distilled water (30-60°C) into the vessel and close the lid.
Press down on the START/ STOP button for 5 seconds, the blue power indicator light, the
yellow low liquid level indicator light and the red low power indicator light will come on.
Switch the device on in cleaning mode to nebulise the distilled water for 1-2 minutes in
order to eliminate the residual medicinal product.
Rinse the vessel and lid with distilled water. Carefully remove any excess water with a
clean, soft cloth or leave the parts to dry naturally in a clean place.
Immerse the medicinal product vessel in a 75% rubbing alcohol solution for at least 3
minutes, then remove the vessel and allow it to dry before storing it in a clean place.
Use medical gauze or cotton buds for cleaning and disinfecting operations.
Do not rub with paper or other cloths, to avoid compromising the correct
operation of the appliance.
Do not clean using sharp objects or touch the mesh.
Do not use household bleach or other aggressive substances to clean the device,
only the solution indicated above is allowed.
Cleaning and sanitising the main unit
Carefully clean the main unit using a 75% rubbing alcohol solution and remove any
excess with a clean, soft cloth.
Clean the electrodes connecting the main unit and the medicinal product vessel using
the procedure indicated above.
Do not clean the main unit using volatile liquids such as petrol or diluents.
Do not rinse the main unit or immerse it in any liquid.
Cleaning and sanitising the mouthpiece and masks
Remove the mouthpiece or mask.
Rinse the mouthpiece and mask with de-mineralised water alone for 2-3 minutes (water
temperature ≤50 °C).
EN
Dry well using medical gauze.
Immerse the mask and mouthpiece in a 75% rubbing alcohol solution for at least 3
30
minutes to disinfect them and leave to dry naturally.
Store the appliance in the carry case provided, when not in use.