CONTROL DISPLAYS AND SYMBOLS
•
All segments displayed (6): Press the ON / OFF button (4) to turn on the
unit; all segments will be shown for 2 seconds.
Memory (7): The last reading will automatically be shown on the display
•
for 2 seconds.
•
Ready for measurement (8): The unit is ready for measurement, the «°C»
icon will keep flashing while the mode icon (ear or forehead) will be displayed.
•
Measurement complete (9): The reading will be shown on the display
3 with the «°C» icon and the mode icon steady. The unit is ready for the
next measurement as soon as the «°C» icon is flashing again.
Surface temperature indication (10): A crossed–ear icon will appear on the
•
display 3 when the reading in the ear mode is outside the range 32.0 – 42.2 °C.
Low battery indicator (11): When the unit is turned on, the battery icon
•
will keep flashing to remind the user to replace the battery.
CHANGING BETWEEN FOREHEAD AND EAR
MODE / SURFACE AND AMBIENT TEMPERATURE
To enable ear measurement, please take the removable cap (1) off the upper
part of the thermometer. The device will immediately switch to ear / surface
and ambient temperature mode. The device is ready for use as soon as the ear
icon appears and the «°C» icon starts flashing.
To change back to the forehead mode, place the removable cap (1) on the ther-
mometer tip and the device will immediately switch back to forehead mode.
The forehead icon will be shown on the display and the «°C» icon will flash to
indicate that the device is ready for measurement.