Applied
12.0
Press.
(mm)
2.0 ATM
10.13
3.0 ATM
10.31
4.0 ATM
10.54
5.0 ATM
10.72
6.0 ATM
10.95
7.0 ATM
11.13
8.0 ATM
11.34
9.0 ATM
11.51
10.0 ATM
11.73
11.0 ATM
11.86
12.0 ATM
12.04
13.0 ATM
12.15
14.0 ATM
12.28
The figures in bold face represent the balloon diameter @ Rated Burst Pressure.
MULLINS-X™ Ultra High Pressure Catheter Sizing Chart
14.0
15.0
(mm)
(mm)
11.69
12.55
11.89
12.81
12.13
13.13
12.37
13.48
12.67
13.89
12.93
14.21
13.27
14.58
13.50
14.80
13.78
15.05
13.93
15.23
14.14
15.46
14.27
15.57
14.43
FOR ALL NuMED CATHETERS AN INFLATION DEVICE
WITH PRESSURE GAUGE SHOULD BE USED.
16.0
18.0
(mm)
(mm)
13.31
15.13
13.72
15.57
14.18
16.09
14.58
16.55
15.03
17.08
15.41
17.40
15.80
17.81
16.02
18.08
16.29
18.41
16.48
18.65
16.71
18.93
4
20.0
22.0
(mm)
(mm)
17.37
20.29
17.74
20.83
18.22
21.46
18.74
21.68
19.27
21.94
19.69
22.20
20.15
22.64
20.50
23.07
20.88
23.46
21.21
23.0
25.0
(mm)
(mm)
19.05
23.07
19.70
23.68
20.48
24.41
21.27
24.69
22.04
24.94
22.71
25.11
23.35
25.51
23.91
25.90