Replacement Items; System Assembly; Attaching Flow Generator; Filling The Chamber - DeVilbiss Healthcare DV5HH Guide D'instructions

Masquer les pouces Voir aussi pour DV5HH:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 19
EN

Replacement Items

Item
Humidification System ....................... DV5HH
Humidifier Chamber ......................... DV5C
Chamber Sealing Gasket ................... DV5C-614

System Assembly

NOtE
- Follow Steps 1-3 if you received your humidifier system and CPAP separately. Start with Step 4 if you received the complete system in one carton.
1.
Remove the air supply tubing from the back of the device.
2.
Locate the air supply port at the bottom of the device. Remove the port plug and insert it into the air supply port in the back of the device (Fig 2).
3.
Locate the heater power connector on the bottom of the device. Remove the connector cover and insert it into the storage opening in the
bottom of the humidifier cradle (Fig 1).

Attaching Flow Generator

4.
For easier attachment, remove the water chamber before assembling the flow generator and humidifier cradle.
5.
Align the base of the device with the cradle at a slight angle as shown (Fig 3). Slide the device forward and align cradle locking tabs with the
recesses in the CPAP . Firmly press down the back to lock into place.
6.
Verify that both sides of the device have engaged into place with the humidifier cradle.
7.
Attach the air supply tubing to the air supply port outlet on the humidifier cradle (Fig 4).
8.
Insert the power cord into the back of the flow generator. Make sure the power cord is fully inserted into the connector (Fig 5). Plug the other
end of the power cord into the AC wall outlet.
cAutION– The height of the system must always be lower than the interface mask during use to prevent water from getting into the CPAP mask.

Filling the chamber

1.
Hold the chamber upright and fill with distilled water to the top of the opening (Fig 6).
cAutION– Use only distilled water that is at room temperature. Do not add any medication or other additives to the water.
2.
Carefully tip the chamber back to the level position. The water in the fill ports will flow into the chamber when it is tipped back to the level
position.
cAutION– For proper operation, ensure the flow generator is OFF before inserting humidifier chamber.
3.
Slide the chamber back into the humidifier cradle. Ensure the chamber is fully engaged with the humidifier cradle and the chamber release latch
snaps in place when inserted.
WARNING
Do not attempt to fill the humidifier while it is attached to the humidifier cradle. Damage to the
humidifier cradle may occur.
If the system is placed on the floor, do not position it where it may be stepped on when getting out of bed.
6
A-DV5HH
Part Number
seT up

Publicité

Table des Matières
loading

Table des Matières