Indications for Use:
For use as an intra-vaginal probe in the treatment of female 'Stress', 'Urge' or
'Mixed' urinary incontinence with muscle stimulation and EMG biofeedback
equipment approved to EN60601-01.
FOR SINGLE PATIENT USE – May be re-used by the same patient.
Use as directed by your medical adviser
Contraindications & Precautions:
DO NOT USE:
With electrical stimulation if fitted with a cardiac pacemaker
•
If the package seal is not intact
surgical scar is present
patients with a history of carcinoma at the site of stimulation
or discomfort occurs and seek medical guidance.
CAUTION IS ADVISED:
When using electrical stimulation on patients with altered or reduced tissue
•
sensation
During electrical stimulation if the user has a history of epilepsy
- seek informed medical guidance
equipment capable of delivering outputs in excess of 2mA/cm
contact the equipment manufacturer or your medical adviser.
Cleaning:
•
Wash PERIFORM
•
Rinse in clean water and dry.
•
Repeat after each use and dry
thoroughly before storing.
•
DO NOT use boiling water.
Disposal:
Clean thoroughly and dispose
of according to local regulations.
Contains nickel:
May cause irritation or allergic reaction following prolonged contact.
Always follow the user instructions.
•
If an infection or tissue damage or recent
•
During menstruation
•
®
+ in warm soapy water.
PERIFORM
+ Instruction Manual - GB
®
•
With electrical stimulation in
When using the PERIFORM
2
•
If pregnant
•
If an irritation
®
+ with
2
. If in doubt
Figure 1