Cleaning / Troubleshooting

Cleaning

Wipe all exterior surfaces with a solution of water and mild detergent after each use.
If contamination occurs, clean according to service manual recommendations.
Important: Always follow your hospital's policy for oxygen compatible cleaning; some cleaners
are highly combustible in the presence of oxygen
WARNINGS
Handle contaminated product in accordance with your hospital's infection
control policy.
Disinfect all suction equipment before shipment to prevent exposure to
hazardous contamination.
CAUTION
Do not steam autoclave or liquid sterilize the venturi/ejector suction module.
Severe impairment to the operation of the venturi/ejector suction module will
result.

Troubleshooting

If the regulator and venturi/ejector suction module do not operate, use the following:
Problem
No suction
Reduced pressure or flow
Important: Refer to the service manual for additional information.
6
6700-0382-000 (Rev. 7) 08/2020
Possible Cause
Venturi is in the O (OFF)
position
Vacuum regulator is OFF
Venturi muffler is dirty or clogged
Bacterial filter clogged
Remedy
Switch to I (ON)
Turn vacuum regulator on
Replace muffler
Replace bacterial filter