Cleaning / Troubleshooting
Cleaning
Wipe all exterior surfaces with a solution of water and mild detergent after each use.
If contamination occurs, clean according to service manual recommendations.
Important: Always follow your hospital's policy for oxygen compatible cleaning; some cleaners
are highly combustible in the presence of oxygen
WARNINGS
Handle contaminated product in accordance with your hospital's infection
⚠
control policy.
Disinfect all suction equipment before shipment to prevent exposure to
⚠
hazardous contamination.
CAUTION
Do not steam autoclave or liquid sterilize the venturi/ejector suction module.
⚠
Severe impairment to the operation of the venturi/ejector suction module will
result.
Troubleshooting
If the regulator and venturi/ejector suction module do not operate, use the following:
Problem
No suction
Reduced pressure or flow
Important: Refer to the service manual for additional information.
6
6700-0382-000 (Rev. 7) 08/2020
Possible Cause
Venturi is in the O (OFF)
position
Vacuum regulator is OFF
Venturi muffler is dirty or clogged
Bacterial filter clogged
Remedy
Switch to I (ON)
Turn vacuum regulator on
Replace muffler
Replace bacterial filter