Reassemble all parts of the nebulizer and connect it to the
air outlet (Pic. G) and switch on the device for 10-15 min-
utes.
Do not boil or autoclave the air hose and the masks.
2. Disinfection:
Cold disinfecting solutions can be used, instead of boiling wa-
ter, according to manufacturer's instructions.
NEBULIZER REPLACEMENT
The nebulizer must be replaced after a long period of inactiv-
ity, in case it shows deformations or breakings, or when the
nebulizer nozzle is obstructed by dry medicine, dust, etc. boso
recommends to replace the nebulizer after a period between
6 months and 1 year, depending on the usage.
!
Use original nebulizers only.
!
The device is only compatible with the boso medisol compact
Accessories kit.

FILTER REPLACEMENT

Under normal usage conditions the air filter must be replaced
every 100 operating hours or every 12 months at the latest.
boso recommends to periodically check the air filter (10 – 12
treatments) and, if the filter shows a grey or brown colour or is
wet, replace it. Remove the filter (Pic. A-4) and replace it with a
new one. Do not attempt to clean the filter to reuse it.
!
Only use original filters.
!
Do not use the device without the filter.