sPeCIFICAtIoNs/
ClAssIFICAtIoNs
speCIfICaTIons/ClassIfICaTIons
Size ...................................................................... 10.1" W x 10.5" H x 6.5" D (26 cm x 27 cm x 17 cm)
Weight ..............................................................................................................................7.1 lbs. (3.2 kg)
Electrical
Requirements 5650D & 5650C ......................................................................1.3 amps, 115VAC 60Hz
5650I ........................................................................................0.8 amps, 230VAC 50Hz
5650K & 5650P ...................................................................0.8 amps, 230VAC 50/60Hz
5650F .......................................................................................0.8 amps, 230VAC 50Hz
Sound Level ...................................................................................................................................51 dBA
Power Consumption 5650D & 5650C ....................................................................90 watts maximum
Maximum Compressor Pressure ...............................................................30 psig or greater (206 kPa)*
Compressor Free Air Flow 5650 ................................................................................. 9 lpm or greater*
Operating Temperature Range all Models except 5650C ..................+40° to +104°F (+5 to +40°C)
Operating Humidity ......................................................................................up to 95% non-condensing
Storage/Transport Temperature Range ................................................ -40° to +140°F (-40° to +60°C)
Storage/Transport Humidity .........................................................................up to 95% non-condensing
Storage/Transport Pressure ....................................................................7.3 to 15.4 psia (50 to 106 kPa)
Supplied Nebulizer Capacity ...................................................................................................... 6 ml (cc)
Supplied Nebulizer Nebulization Rate ............................................................... 0.15 ml/min or greater*
Supplied Nebulizer MMAD ....................................................................................... 5 microns or less**
*Conditions may vary based on altitude above sea level, barometric pressure, and
temperature.
**Value determined with minimum flow rate through the nebulizer (20°C, sea level).
5650F Classifications
With respect to protection against electric shock ........................................................................Class II
Degree of protection against electrical shock .......................................................Type BF Applied Parts
Degree of protection against ingress of liquids ............................................................................... IPXO
Mode of Operation ............................................................................................................... Continuous
5650I, 5650K & 5650P ........................................................100 watts maximum
5650F ...................................................................................105 watts maximum
5650C ............................................+40° to +101°F (+5 to +38.5°C)